Internal dosimetry study of [Rb-82]Cl using a long axial field-of-view PET/CT

Mercolli, L., Bregenzer, C., Diemling, M. et al. Internal dosimetry study of [82Rb]Cl using a long axial field-of-view PET/CT. Eur J Nucl Med Mol Imaging (2024).

https://doi.org/10.1007/s00259-024-06660-7

Abstract

Purpose

Long axial field-of-view (LAFOV) positron emission tomography (PET) systems allow to image all major organs with one bed position, which is particularly useful for acquiring whole-body dynamic data using short-lived radioisotopes like Rb-82.

Methods

We determined the absorbed dose in target organs of three subjects (29, 40, and 57 years old) using two different methods, i.e., MIRD and voxel dosimetry. The subjects were injected with 407.0 to 419.61 MBq of [82Rb]Cl and were scanned dynamically for 7 min with a LAFOV PET/CT scanner.

Results

Using the MIRD formalism and voxel dosimetry, the absorbed dose ranged from 1.84 to 2.78 μGy/MBq (1.57 to 3.92 μGy/MBq for voxel dosimetry) for the heart wall, 2.76 to 5.73 μGy/MBq (3.22 to 5.37 μGy/MBq for voxel dosimetry) for the kidneys, and 0.94 to 1.88 μGy/MBq (0.98 to 1.92 μGy/MBq for voxel dosimetry) for the lungs. The total body effective dose lied between 0.50 and 0.76 μSv/MBq.

Conclusion

Our study suggests that the radiation dose associated with [82Rb]Cl PET/CT can be assessed by means of dynamic LAFOV PET and that it is lower compared to literature values.

Introduction

Due to its cost-effectiveness, [82Rb]Cl is increasingly used in positron emission tomography (PET) for myocardial perfusion imaging (MPI) [1,2]. With a broader clinical adoption, it is important to have accurate and reliable estimates for the radiation dose that is delivered to the patient. However, only a few studies investigated dosimetry with [82Rb]Cl PET/CT, and wide discrepancies can be seen in the reported values [3,4,5,6,7,8].

Previous biokinetic and dosimetric studies for [82Rb]Cl are plagued by the short half-life of the radiotracer (76 s) and by the use of analogue PET/CT with a standard axial field-of-view (SAFOV). Moreover, while it is theoretically possible to use blood flow as a surrogate quantity to estimate the biokinetics of [82Rb]Cl [3,8], the quantification of the strong model dependence is challenging. Note also that SAFOV PET/CT protocols require multiple bed positions and therefore longer scan times to image all relevant organs [4,5,7] and a whole-body dynamic acquisition is not feasible [9]. The introduction of long axial field-of-view (LAFOV) PET/CT systems in clinical routine [10,11] enables whole-body dynamic imaging and kinetic modelling with unprecedented accuracy [12,13]. This allows in turn for an accurate estimate of the absorbed radiation dose.

The aim of this study is to increase the accuracy of the dose estimates for [82Rb]Cl imaging, which were to date hampered by the use of surrogate quantities (blood flow) or large injected activities (to compensate for long scan times). To that end, we estimated the absorbed doses in healthy volunteers, by acquiring dynamic [82Rb]Cl PET images on a LAFOV PET/CT scanner.

Materials and methods

Subjects

Three healthy volunteers were scanned first at rest and immediately after under stress conditions. Pregnancy was excluded at the time of acquisition and none had diffuse atherosclerosis or coronary artery disease (CAD). The pharmacological stress was induced with 400 mcg of Regadenoson. The injected [82Rb]Cl activity ranged from 407.0 to 419.61 MBq. The subjects’ characteristics and the administered activities of [82Rb]Cl are displayed in Table 1. [82Rb]Cl was produced with CardioGen-82 radionuclide generator and infusion system (Bracco Imaging S.p.A, Milan, Italy). The radiotracer was automatically infused in an antecubital vein of the left arm over 20 s. In Fig. 1, the maximum intensity projections (MIP) images of the 3 subjects are shown.

Nuclear medicine scans. Maximum intensity projection (MIP) of the three subjects at rest. The images depict the full scan integrationFigure1: Maximum intensity projection (MIP) of the three subjects at rest. The images depict the full scan integration.Maximum intensity projection (MIP) of the three subjects at rest. The images depict the full scan integration

Imaging protocol

Images were acquired on a LAFOV PET/CT scanner (Biograph Vision Quadra, Siemens Healthineers, Knoxville, TN, USA) [14]. The scans started simultaneously with the administration of [82Rb]Cl and electrocardiogram-ungated images were acquired in list mode over 7 min both at rest and under stress.

The images were reconstructed using a dedicated image reconstruction prototype (e7-tools, Siemens Healthineers) to a 440 × 440 matrix, with 3-mm slice thickness, 4 iterations, 5 subsets, time-of-flight (TOF), point-spread-function (PSF) recovery, and a 2-mm full width at half maximum Gaussian filter. Seven time points were reconstructed for each scan, i.e., integrating the accumulated counts between 0 and 30, 30 and 60, 60 and 90, 90 and 120, 120 and 180, 180 and 270, and 270 and 420 s, respectively. Each image was reconstructed using a proprietary ordered-subset expectation maximum (OSEM) iterative algorithm. The images were corrected for attenuation and scatter based on a low-dose CT (dose-length product 27.4 to 41.6 mGy cm).

Dosimetry

We retrospectively determined the time-activity curves (TACs) from the LAFOV PET images of each subject. To this end, 11 source organs (adrenals, gallbladder, stomach, heart wall/ventricular cavities, kidneys, liver, lung, spleen, thyroid, urinary bladder, uterus) were segmented using the corresponding CT images and applied to all time points in the PET images. The organ segmentation was performed using the artificial intelligence–based TotalSegmentator tool [15] and an experienced nuclear medicine physician verified the automatic segmentation as needed (the segmentation is shown in Supplemental Fig. 1). For the TACs, the mean intensity value has been decay corrected for the decay during each frame and to the beginning of the scan.

To compute the dose based on the MIRD formalism [16], we applied a double exponential fitting function to the seven time points of the TACs and removed the decay correction to the injected activity (hence only the decay within a time frame is corrected). The fitted function was integrated over time in order to obtain the time-integrated activity (TIA) for each source organ. We used the subjects’ TIA as input for Olinda/EXM version 2.2.3 (Hermes Medical Solutions, Stockholm, Sweden) to compute the organ doses for the ICRP adult male and female reference phantoms [17].

As a second method, we used a prototype voxel dosimetry software (Hermes Medical Solutions, Stockholm, Sweden) to compute a fully individual patient dose. The voxel-wise TIA is computed through a trapezoidal time integration over the seven time points and a mono-exponential decay after the last time point. The decay parameter of the mono-exponential decay is fitted from the last two time points. The same organ segmentation of the PET images is applied to the resulting dose maps to obtain the organ doses.

Results

In Fig. 2, we show the TACs of selected organs at rest and under stress. These values were obtained by integrating a double exponential fitting function over time. All organs show a rapid increase within the first minute, which stabilizes after about 2 min. In Table 2, we report the TIA for the three subjects at rest and under stress.

Nuclear medicine graph of Decay corrected TAC of selected organsFigure2: Decay corrected TAC of selected organs.

In Table 3, we report the normalized organ doses as obtained from Olinda/EXM. For the most relevant organs, we visualize the absorbed doses in Fig. 3, where we added also the organ doses computed from Hermes’ voxel dosimetry module (see Table 4 for the full voxel dosimetry results).

Nuclear medicine graph of normalized absorbed dose for selected organs as obtained from Olinda/EXM and voxel dosimetry. The dashed line indicates the corresponding dose from the ICRP.Figure3: Normalized absorbed dose for selected organs as obtained from Olinda/EXM and voxel dosimetry. The dashed line indicates the corresponding dose from the ICRP.

The total body effective dose is reported in Table 5 and lies between 0.50 and 0.76 μSv/MBq. As expected from the TIA in Table 2, the effective doses for the three subjects are consistently lower than the reference value from the ICRP 128 publication (also depicted in Supplemental Fig. 2) [8].

Discussion

The unique possibility to image all main organs of a patient with a single bed position allows [82Rb]Cl LAFOV PET/CT to go beyond the standard paradigms of cardiovascular imaging in patients with suspected or known coronary artery disease (CAD) [9,18,19]. Due to the significant increase in sensitivity compared to SAFOV systems, the administered activity can be kept to a minimum [10,18] and this advantage reflects our choice to administer activities below the to-date recommended range (740 to 1480 MBq) [2]. Hence, the injected activities in our study are significantly lower than those administered in the previous reports [4,5,7]. This is also consistent with the manufacturer’s recommendation, suggesting that activities above 555 MBq [82Rb]Cl would likely saturate the scanner’s data acquisition.

We found that almost all TIA are smaller than reported in ICRP publication 128 [8]. This is in line with the conservative nature of the ICRP values and they do not represent individual estimates for a single patient. Furthermore, the strong model dependence of [8] conceivably drives the TIA to higher values. Furthermore, our TIA is slightly lower than reported in [4,7]. While it should be noted that the TAC and TIA vary among the three subjects, the same issue also pertains to the abovementioned reports, wherein the standard deviation across patients in the cohort was quite large.

Consistent with previous studies, we found differences in the TACs between rest and stress acquisitions. Specifically, while some organs show an increased uptake on stress (e.g., heart, liver), others present with higher TACs on rest (e.g., kidneys, spleen). Comparing the rest and stress TIA, our results are consistent with the reports from Senthamizhchelvan et al. and Mattssonn et al. [5,7], wherein similar differences were seen.

For some target organs, the normalized dose obtained from Olinda/EXM showed an excess compared to the ICRP 128 values (see Table 3 and Fig. 3). Given that TIA in our study is consistently lower than in [8], the most conceivable explanation lies in the differences in human phantoms (and therefore in the S-values) that underlie the ICRP 128 and the Olinda/EXM dose calculation. Furthermore, the normalized total body effective doses shown in Table 2 are smaller than the reports of [7,8]. Interestingly, [4] report a total body effective dose, which is even slightly higher than [8] and almost double compared to our results. The contention may relate to either the rather large systematic uncertainties in the dose estimation in general or to a possible overestimation of the TIA in [4] due to their fitting procedure that uses only three time points.

On average, the difference between rest and stress in the absorbed dose is marginally lower for the three subjects. For the gallbladder wall, heart wall (only subjects P1 and P2), kidneys (only subjects P1 and P3), liver, lungs, spleen (only subject P1), thyroid and urinary bladder (only subject P1), the relative difference between rest and stress dose exceeds 5%. The resulting total body effective dose reflects the results from the organ doses, i.e., the relative differences are 1.6, − 9.2, and − 0.7% for the subjects P1, P2, and P3, respectively. Given the small difference, a pharmacological-induced stress is unlikely to affect patient’s dose, which is in line with the ICRP’s assumption to disregard the effects of physiological rest and stress on absorbed dose to the patient.

The organ doses from the voxel dosimetry show an overall consistency with the doses from Olinda/EXM. We attribute the deviations seen in Fig. 3 to intrasubject variabilities and the uncertainties in the TIA determination at the voxel as well as at the organ level. This is in line withs previous comparisons between Olinda/EXM and Hermes’ voxel dosimetry [20,21], albeit in the context of PET tracer rather than therapeutic applications.

The main limitation of our study is the limited number of subjects, due to the current unavailability of cardiac gating for the LAFOF-PET scanner, which forbids the use for clinical routine. Despite the small patients’ cohort, our results provide a strong rationale to further investigate the role of LAFOV PET in dosimetry studies with [82Rb]. Moreover, our work suggests that a reassessment Cl dosimetry should be pursued, taking advantage of the possibility to perform “whole-body” dynamic imaging [9]. Second, we did not correct our data for partial volume effect (PVE), and it may be argued that a model of biodistribution may suffer from PVE depending on how it was derived. Indeed, a more pronounced difference should be seen in the voxel dosimetry results, given the fact that the dose is inversely proportional to the volume. However, TIA of the segmented organs should not be significantly affected by PVE, since it is derived from the mean voxel values of the segmented organs. Considering mean voxel values, TACs in our study show consistency across patients. Hence, it is conceivable that the impact of PVE is not relevant.

Summarizing, our study shows that LAFOV PET/CT can be used to estimate absorbed dose from [82Rb]Cl imaging. Our results hint towards lower values of the normalized absorbed dose compared to the literature, while the use of LAFOV PET/CT also enables the use of very low-activity protocols. This gives more reliance in suggesting [82Rb]Cl PET/CT in clinical practice with acceptable radiation exposure. Similarly, to previous reports, TACs and TIA varied to a certain extent among the three subjects; hence, inter-subject discrepancies should be considered.

Conclusion(s)

[82Rb]Cl internal dosimetry plays a key role in the overall assessment and optimization of MPI. Our study provides absorbed dose estimates at rest and under pharmacological stress, showing lower values than currently reported in the literature. Hence, it should be of less concern to use [82Rb]Cl PET in the assessment of patients with suspected or known coronary artery disease. In addition, the increased sensitivity of LAFOV PET/CT systems allows for a significant reduction of the administered activity, thus allowing a further reduction of the absorbed dose compared to standard SAFOV PET/CT scanners. 

References 

  1. Mimouni M, Bulsei J, Darlington M, Estellat C, Rouzet F, Hyafil F, et al. Cost-effectiveness of 82-Rubidium PET myocardial perfusion imaging for the diagnosis of myocardial ischemia depending on the prevalence of coronary artery disease. EJNMMI Res. 2023;13:9.

  2. Sciagrà R, Lubberink M, Hyafil F, Saraste A, Slart RHJA, Agostini D, et al. EANM procedural guidelines for PET/CT quantitative myocardial perfusion imaging. Eur J Nucl Med Mol Imaging. 2021;48:1040–69.

  3. Leggett RW, Williams LR. A proposed blood circulation model for reference man. Health Phys. 1995;69:187–201.

  4. Senthamizhchelvan S, Bravo PE, Esaias C, Lodge MA, Merrill J, Hobbs RF, et al. Human biodistribution and radiation dosimetry of 82Rb. J Nucl Med. 2010;51:1592–9.

  5. Senthamizhchelvan S, Bravo PE, Lodge MA, Merrill J, Bengel FM, Sgouros G. Radiation dosimetry of 82Rb in humans under pharmacologic stress. J Nucl Med. 2011;52:485–91.

  6. Stabin MG. Proposed revision to the radiation dosimetry of 82Rb. Health Phys. 2010;99:811–3.

  7. Hunter CRRN, Hill J, Ziadi MC, Beanlands RSB, deKemp RA. Biodistribution and radiation dosimetry of (82)Rb at rest and during peak pharmacological stress in patients referred for myocardial perfusion imaging. Eur J Nucl Med Mol Imaging. 2015;42:1032–42.

  8. Mattsson S, Johansson L, LeideSvegborn S, Liniecki J, Noßke D, Riklund KÅ, et al. Radiation dose to patients from radiopharmaceuticals: a compendium of current information related to frequently used substances. Ann ICRP. 2015;44:7–321.

  9. Caobelli F, Gräni C, Rominger A. The future of atherosclerosis assessment: dynamic and quantitative whole-body PET cardiovascular imaging. Eur J Nucl Med Mol Imaging. 2023;50(11):3177–80https://doi.org/10.1007/s00259-023-06292-3.

  10. Alberts I, Hünermund J-N, Prenosil G, Mingels C, Bohn KP, Viscione M, et al. Clinical performance of long axial field of view PET/CT: a head-to-head intra-individual comparison of the biograph vision quadra with the biograph vision PET/CT. Eur J Nucl Med Mol Imaging. 2021;48:2395–404.

  11. Slart RHJA, Rominger A. Collection on clinical total-body-PET studies. Eur J Nucl Med Mol Imaging. 2023;50:1556–7.

  12. Sari H, Mingels C, Alberts I, Hu J, Buesser D, Shah V, et al. First results on kinetic modelling and parametric imaging of dynamic 18F-FDG datasets from a long axial FOV PET scanner in oncological patients. Eur J Nucl Med Mol Imaging. 2022;49:1997–2009.

  13. Yu X, Sun H, Xu L, Han Y, Wang C, Li L, et al. Improved accuracy of the biodistribution and internal radiation dosimetry of 13 N-ammonia using a total-body PET/CT scanner. Med Phys. 2023;50(9):5865–74. https://doi.org/10.1002/mp.16450.

  14. Prenosil GA, Sari H, Fürstner M, Afshar-Oromieh A, Shi K, Rominger A, et al. Performance characteristics of the Biograph Vision Quadra PET/CT system with a long axial field of view using the NEMA NU 2–2018 standard. J Nucl Med. 2022;63:476–84. Available from: https://jnm.snmjournals.org/content/63/3/476

  15. Wasserthal J, Breit HC, Meyer MT, Pradella M, Hinck D, Sauter AW, et al. TotalSegmentator: Robust Segmentation of 104 Anatomic Structures in CT Images. Radiol Artif Intell. 2023;5(5):e230024. https://doi.org/10.1148/ryai.230024.

  16. Bolch WE, Eckerman KF, Sgouros G, Thomas SR. MIRD pamphlet No. 21: a generalized schema for radiopharmaceutical dosimetry–standardization of nomenclature. J Nucl Med. 2009;50:477–84.

  17. International Commission on Radiological Protection. Basic anatomical and physiological data for use in radiological protection: reference values. ICRP Publication 89. Ann ICRP. 2002;32:5–265.

  18. Caobelli F, Seibel S, Krieger K, Bregenzer C, Viscione M, Silva Mendes AF, et al. First-time rest-stress dynamic whole-body 82Rb-PET imaging using a long axial field-of-view PET/CT scanner. Eur J Nucl Med Mol Imaging. 2023;50:2219–21.

  19. Mingels C, Sari H, Gözlügöl N, Bregenzer C, Knappe L, Krieger K, et al. Long-axial field-of-view PET/CT for the assessment of inflammation in calcified coronary artery plaques with [68 Ga]Ga-DOTA-TOC. Eur J Nucl Med Mol Imaging. 2024;51(2):422–33https://doi.org/10.1007/s00259-023-06435-6.

  20. Hippeläinen ET, Tenhunen MJ, Mäenpää HO, Heikkonen JJ, Sohlberg AO. Dosimetry software Hermes Internal Radiation Dosimetry: from quantitative image reconstruction to voxel-level absorbed dose distribution. Nucl Med Commun. 2017;38:357–65.

  21. Willowson KP, Ryu H, Jackson P, Singh A, Eslick E, Bailey DL. A comparison of 2D and 3D kidney absorbed dose measures in patients receiving 177Lu-DOTATATE. Asia Ocean J Nucl Med Biol. 2018;6:113–9.

 

Additional information

This article is licensed under a Creative Commons Attribution 4.0